Our Products
   
By Product Type, Brand or Keyword
Services

Featured Products & Specials
AMI Biotech Parts
Parts, Accessories, Rental, Repair
Request a Quote / More Information
Customer Survey
Operations Manuals - Download
Ardus Medical
Medical Supply, Hospira, Abbott, Infusion Pump, Medical Equipment - Ardus Medical

Infusion Pumps

Abbot Plum A+ Infusion Pump

Hospira/Abbott Plum A+ Infusion Pump

Save up to 80% on pre-owned Hospira/Abbott Labs Plum A+ infusion pumps.
Request a Quote for this product now.


Product Specifications

Dimensions:
8" H x 8" W x 6" D, excluding pole clamp

Weight:
9.5 lbs. with battery

Casing:
High-impact plastic

Delivery Rate Range

Lines A & B:
0.1-99.9 ml/hr in 0.1 ml increments
100-999 ml/hr in 1 ml increments

Concurrent Delivery:
0.5 ml/hr minimum for each line

Plum Set:
500 ml/hr cumulative (A+B) maximum

KVO:
1.0 ml/hr or the last primary delivery rate, whichever is less

VTBI Range:
0.1-99.9 ml in 0.1 ml increments
100-9999 ml in 1 ml increments

Electrical

Power Requirements:
120V, 50-60 Hz, 35VA

Fuses:
F1, F2, 250V, 0.5A

Power Cord:
Hospital Grade AC Cord 10 ft long

Battery:
One sealed lead acid, rechargeable 6V battery, internal

Battery Life:
With a new fully charged battery, the infuser shall operate for a minimum of 6 hours at 125 ml/hr or less, or deliver 500 ml at 126 ml/hr or greater on one line

Recharge:
The battery charges whenever the pump is connected to AC power.  The recharge time is approximately 6 hours with the device operating at 125 ml/hr on one line

Copyright © 2010 All Rights Reserved. Ardus Medical, Inc. | Site by Justice & Young | 1.800.878.1388 | info@ardusmedical.com | Site Map