Our Products
   
By Product Type, Brand or Keyword
Services

Featured Products & Specials
AMI Biotech Parts
Parts, Accessories, Rental, Repair
Request a Quote / More Information
Customer Survey
Operations Manuals - Download
Ardus Medical
Alaris, Medical Equipment, Infusion Pump Service, Infusion Pump - Ardus Medical

Infusion Pumps

Alaris/Cardinal Medley System Infusion Pump

Alaris/Cardinal Medley System Infusion Pump

Service only, not a product for sale/rent.


Product Specifications

Point of Care Unit (POC)

Dimensions:
6.9"W x 8.8"H x 9"D including pole clamp

Weight:
7.2 lbs.

Battery Operation:
Battery run time is a function of the number of modules attached and module activity. See operator's manual for battery operation hours.

Communication Data Port:
RS-232 with an RJ45 connector

Electronic Memory:
System configuration parameters stored in volatile memory are retained for at least 6 months by an internal backup lithium battery.  Module specific parameters are stored for 8 hours when system is turned off.

Power Requirements:
100-240V, 50/60 Hz, 150VA Max

Pump Module

Dimensions:
3.3"W x 8.9"H x 5.5"D

Weight:
2.5 lbs.

Operating Principle:
Positive Displacement

Flow Rate:
0.1-99.9 ml/hr in 0.1 ml increments
100-999 ml/hr in 1 ml increments

Volume to be Infused:
0.1-9.99 ml in 0.01 ml increments
10-999.9 ml in 0.1 ml increments
1000-9999 ml in 1 ml increments

KVO Rate:
1ml/hr or selected rate, whichever is lower

Infusion Pressure Maximum:
654 mmHg

Rate Accuracy:
±5% at rates between 1 and 999 ml/hr
±5.5% at rates less than 1 ml/hr

Syringe Module

Dimensions:
4.5"W x 15.0"H x 7.5"D

Weight:
4.5 lbs

Operating Principle:
Positive Displacement

Flow Rate:
0.01-9.99 ml/hr in 0.01 ml increments
10-99.9 ml/hr in 0.1 ml increments
100-999 ml/hr in 1 ml increments

Volume to be Infused:
0.1-9.9 ml in 0.01 ml increments
10-60 ml in 0.1 ml increments

Rate Restriction by Syringe Size:
50-60 ml syringe - 0.1-999 ml/hr
30 ml syringe - 0.1-650 ml/hr
20 ml syringe - 0.1-500 ml/hr
10 ml syringe - 0.1-250 ml/hr
5 ml syringe - 0.1-150 ml/hr
3 ml syringe - 0.01-100 ml/hr
1 ml syringe - 0.01-30 ml/hr

KVO Rate:
1ml/hr or selected rate, whichever is lower

Infusion Pressure Maximum:
800 mmHg without pressure sensing disc
1060 mmHg with pressure sensing disc

Rate Accuracy:
±2% of full scale plunger travel not including syringe variation

PCA Module

Dimensions:
4.5"W x 15.0"H x 7.5"D exclusive of security door

Weight:
5.5 lbs.

Operating Principle:
Positive Displacement

Delivery Units:
mg, mcg, mL, mg/h, mcg/h, mL/h

Flow Rate:
0.10-9.99 ml/hr in 0.01 ml increments
10-99.9 ml/hr in 0.1 ml increments
100-999 ml/hr in 1 ml increments

Rate Restriction by Syringe Size:
50-60 ml syringe - 0.1-999 ml/hr
30-35 ml syringe - 0.1-650 ml/hr
20 ml syringe - 0.1-500 ml/hr

Occlusion Alarm Thresholds:
Low, Medium, and High

Rate Accuracy:
±2% of full scale plunger travel not including syringe variation

Nellcor® SPO2 Module

Dimensions:
3.3"W x 8.9"H x 5.5"D

Weight:
2 lbs.

Mode of Operation:
Continuous

Measurement Range:
Perfusion - 0.03 to 20%
Pulse Rate - 20-250 bpm
SPO2 - 1-100%

Sensor:
Emitted light wavelength range is within 500 nm to 1000nm