Cadd 5700 TPN Ambulatory Pump
Save up to 80% on pre-owned Cadd 5700 TPN ambulatory pumps.
Request a Quote for this product now.
Product Specifications
- Pump Resolution:
- 250 microliters per pump stroke (nominal)
- Dimensions:
- 1.1" D x 3.5" W x 5.25"H including set plate
- Weight:
- 13 oz including 9 volt battery
- Power Source:
- 9 volt alkaline or lithium battery, or External Power Source
- Pump Timing Accuracy:
- ±0.1%
- System Delivery Accuracy:
- ±8% nominal
- Reservoir residual volume:
- 0-9990 ml, in 10 ml increments
- Infusion volume:
- 10-9990 ml, in 10 ml increments
- Infusion Period:
- 10 minutes to 99 hours and 50 minutes (99:50), in 10 minute increments. The pump will control variations in the programming schedule to maintain the continuous rate within the range of 10-400 ml/hr.
- Taper Up Period:
- 0-99 hours and 40 minutes (99:40) in 10 minute increments
- Taper Down Period:
- 0-99 hours and 40 minutes (99:40) in 10 minute increments
- Continuous Infusion Rate:
- 10-250 ml/hr with a 9 volt battery
- 10-400 ml/hr with the power pack and/or AC Adapter
- The rate is established by the programmed values of infusion volume, infusion period, and tapering (if any).
- Delivered Infusion Volume:
- 0-9990 ml, in 10 ml increments
- Optional KVO infusion rate:
- 5 ml/hr
- Pump Alarms:
- Low battery power
- pump in Stop mode
- internal controller, microprocessor, or pump motor fault
- improper delivery
- power up fault
- low fluid container volume
- high delivery pressure (22 ± 12 psi)
|