Ambulatory Pump, Medical Supply, Medical Equipment - Ardus Medical
Anesthesia
Drager: Narkomed GS
Product Specifications
- Features:
- • Weight (base unit without vaporizers or cylinders): 224 lbs (101.6 kg)
- • Dimensions: (W) 89.5 cm x (H) 130 cm x (D) 82 cm 35.2 x 51.2 x 32.3 inches)
- • Power Supply: 100-240 VAC, 50/60 Hz, 2.3 A max.
- • Battery (supports ventilator and monitor): > 45 min
- • Ventilator E-vent: Electronically controlled, electrically driven
- • Breathing Frequency: 4 to 60 bpm
- • Max. Minute Volume (MV): 25 L/min
- • Positive End-Expiratory Pressure (PEEP): 0-20 cmH2O
- • Inspiration/Expiration Ratio (Ti:Te): 4:1 to 1:4
- • Pressure Limiting (Pmax): 15-70 cmH2O
- • Tidal Volume (Vt): 20-1400 mL in Volume Control; 20-1100 mL in SIMV/PS
- • Inspiratory pause (Tip:Ti): 0-50%
- • SIMV Inspiratory Time (Tinsp): 0.3-4.0 sec
- • Inspiratory Pressure (Pinsp): PEEP + 5 to 65 cmH2O
- • Inspiratory Flow (InspFlow): 10-75 L/min in Volume and Pressure Control, 10-85 L/min in Pressure Support
- • Pressure Support Level (ΔPPS): PEEP + 3 to 20 cmH2O
- • Min. Frequency for Apnea-Ventilation (Freq. Min.): 3-20 bpm and "OFF"
- • Trigger: 2-15 L/min
- • Range of Fresh Gas Flow Indicators: 0.00 to 12.0 L/min
- • Total Fresh Gas Flow Meter: 0 to 10 L/min, calibrated with a mixture of 50 % O2 and 50 % N2O mixture
- • O2 Flush: at 87 psi (6 bar): max.75 L/min, at 41 psi (2.8 bar): min 25 L/min
- • Vaporizer Mount: Drager or SelectatecR mount
- • Serial Interface: 1 x RS 232 (standard), 1 x RS 232 (option)
- • Protocols: Vitalink and Medibus
- • Data Available for Export: All fresh gas flow, ventilation and O2 data
- • Volume of CO2 Absorber: 1.5 Liter, option: Drager medical's consumable CLIC adsorber
- • Volume of Entire compact Breathing System: 2.8 Liter + bag
|